    Raycell® Mk1
Enquiry Now
Technical Specifications :
| Weight | Photon Energy | 
|---|---|
| 550 kg (1200 lbs) | 160 keV max, 60-80 keV average | 
| Floor Loading | Cooling System | 
| 70 kg/cm2 (1000 lbs/in2) | Self-contained cooling system — no external water required | 
| Electrical Requirements | Environmental Requirements | 
| Single-phase 60Hz, 21A typically at 208V | Temperature: 10°-30°C (50°-86°F); Relative Humidity: 5%-75% | 
| Single-phase 50Hz, 19A typically at 230V | |
| Dimensions | |
| Height | 1255 mm (49.4 in) | 
| Width | 1060 mm (41.75 in) | 
| Depth with UPS | 975 mm (38.4 in) | 
| Depth without UPS | 870 mm (34.25 in) | 
Canister Features :
| Volume | 1.7 L | 
| Diameter | 173 mm (6.8 in) | 
| Height | 79 mm (3.1 in) | 
| Estimated Number of Bags per Cycle | 3 Bags, 300 ml each | 
| Central Dose Rate | ~6.3 Gy/min | 
| Time for 25 Gy Central Dose~4 min | ~4 min | 
| Dose Uniformity | 19.6 to 34.3 Gy (1.75:1) | 
| Time for 25 Gy Minimum Dose | ~5 min | 
| Dose Uniformity | 25 to 43.8 Gy (1.75:1) | 
Options & Accessories :
- Barcode Reader
- Syringe Holder
- LIMS Connection
- Printer Connection
- Additional Canisters
- Irradiation Verification Indicators
- Service Agreements
- Dosimetry
- UPS